IsoMist - Glass Expansion
Download
Report
Transcript IsoMist - Glass Expansion
A Programmable Universal
Temperature Controlled
Spray Chamber for ICPOES and ICP-MS
Jerry Dulude, Ron Stux, and Vesna Dolic, Glass
Expansion, jdulude@geicp.com
Limitations of Room Temperature
Spray Chambers
Sensitivity drift with temperature
Excessive plasma loading (volatile solvents)
Excessive oxide formation
Insufficient control of analyte transport
Glass Expansion
www.geicp.com
1
Limitations of Externally Controlled
Spray Chambers
Requires antifreeze solution
Requires bulky floor model chiller
Cumbersome coolant lines from chiller to
chamber
Freezing of condensate on lines
Requires intricate jacketed chamber
Glass Expansion
www.geicp.com
1
IsoMist™ Controlled Temperature
Spray Chamber
Glass Expansion
www.geicp.com
1
IsoMist Encapsulated Spray Chamber
Glass Expansion
www.geicp.com
1
Features of IsoMist
Powerful Peltier-effect chiller reaches -5C in 15
minutes (chamber interior)
Programmable from -10 to 60C in 1 degree
increments
Maintains temperature to within 0.5 degree
Compact design (7.5x4x4 inches)
100% self-contained (no external lines)
Incorporates Bluetooth® technology for clean
wireless control (USB available)
Compatible with all ICP-OES and ICP-MS models
Glass Expansion
www.geicp.com
1
PC Screen showing IsoMist Software
Glass Expansion
www.geicp.com
1
Determination of Trace Metals in
Naphtha
Interfere with the cracking process
Poisoning of the catalysts
Environmental release concerns
Origin and migration markers
Too volatile for room temperature analysis due to
excessive plasma loading
Glass Expansion
www.geicp.com
1
Analysis of Naphtha by ICP-OES
PE Optima 2100 DV (Axial view)
Forward Power: 1500Watts
Coolant flow: 20L/min
Auxiliary flow: 1.8L/min
Nebulizer gas flow: 0.35L/min
Injector: 1mm capillary bore
Spray Chamber: Twister baffled cyclonic
Nebulizer: SeaSpray glass concentric
Uptake rate: 300ul/min
Glass Expansion
www.geicp.com
1
IsoMist on Optima 2100DV
Glass Expansion
www.geicp.com
1
Analysis of Naphtha by ICP-OES
Sample Preparation
Blanks & Standards: 100% kerosene
Samples: 100% Naphtha
Spikes: Naphtha + S-21 (Conostan)
Internal Standard: 0.5ppm Co (Conostan)
IsoMist Temperature: -10C
Glass Expansion
www.geicp.com
1
Naphtha Reproducibility
Neat Sample
Concentration(ppm)
Reproducibility of naphtha analysis (90 min apart)
1.200
1.100
1.000
0.900
0.800
0.700
0.600
0.500
0.400
0.300
0.200
0.100
0.000
Cd 228.802
Cr 205.560
Cu 324.752
Fe 259.939
Mn 257.610
Ni 341.476
P 214.914
Pb 220.353
1
2
Sn 189.927
Ti 334.940
naphtha(ppm)
V 292.464
analysis
Glass Expansion
www.geicp.com
1
Spike Recoveries
800%
700%
600%
500%
400%
0.1 spike recovery
300%
200%
100%
0%
0
9
6
2
2
0
4
40
14
53
27
9
80
75
56
61
93
47
9
46
3
9
.
.
.
.
.
.
.
.
.
.
.
4
8
4
5
7
9
1
4
2
0
9
33
25
25
34
22
32
22
18
20
21
29
i
i
r
n
P
T
V
N
C
Fe
Cd
Cu
Pb
Sn
M
Glass Expansion
www.geicp.com
1
Analysis of Limited Sample Volumes
Biological applications
Intracellular fluids
Neonatal samples
Forensic applications
Requires micro flow nebulization
Low DL’s desirable
Alternative to HEN/Direct injection
Glass Expansion
www.geicp.com
1
39
4
Cd .40
21 1
4
Cd .44
22 0
Co 6.50
23 2
8
Cr .89
26 2
7
Cr .71
20 6
5
Cu .56
32 0
Cu 7.39
32 3
4
Fe .75
23 2
8
Ni .20
23 4
1
Ni .6 0
22 4
1
Pb .6 4
22 8
0
Se .35
19 3
6.
02
V
29 6
2
Zn .46
20 4
6
Zn .20
21 0
3
Zn .85
20 7
2.
54
8
Al
Relative Sensitivity
Effect of IsoMist Temperature on
Normalized Sensitivity (20ul/min Uptake)
4.00
3.50
3.00
2.50
Glass Expansion
40C
2.00
60C
1.50
21C
1.00
0.50
0.00
PE Optima 2100DV
www.geicp.com
1
9
Al 6.1
39 53
Be 4.
4
31 01
C 3.1
d
2 07
C 14.4
d
22 40
C 6 .5
r2 0
6 2
C 7 .7
r2 1
6
C 05.
u
56
3
0
C 27.3
u
32 93
F e 4 .7
2 52
M 38.
2
n
25 04
N 7 .6
i 2 10
3
N 1.6
i2 0
2 4
Pb 1.6
22 48
Se 0.3
19 53
V 6.02
29 6
Z n 2 .4
20 64
Z n 6 .2
21 00
Z n 3 .8
20 57
2.
54
8
Al
3
Rel. DL
Effect of IsoMist Temperature on DL
12
10
8
2ml/min 21C
6
Glass Expansion
.02ml/min 21C
.02ml/min 60C
4
2
0
PE Optima 2100DV
www.geicp.com
1
IsoMist Benefits
Enables the analysis of volatile organics
Enhances sensitivity for limited volume
samples
Reduces isobaric oxide interferences
Increases the chance of passing QC’s
Provides a record for regulatory compliance
2C = 10% shift in sensitivity
Glass Expansion
www.geicp.com
1